| Model Number |
1012272-20 |
| Medical Device Problem Codes |
Material Separation (1562); Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that when opening and removing the 2.5x20mm trek balloon dilatation catheter (bdc) from it's packaging the plastic loop was kinked and the bdc had a proximal shaft separation.Another device was used to successfully complete the procedure.There was no patient involvement and no clinically significant delay.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection was performed on the returned device.The reported kinked coil and separation were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation was unable to determine a conclusive cause for the reported kinked coil; however, the separation appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|