|
MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOU
|
Back to Search Results |
|
| Model Number |
DM0010FAA |
| Medical Device Problem Code |
Device Remains Activated (1525)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
08/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No conclusion can be drawn.No evaluation was performed, as the device was discarded by the customer.This dm0010faa easydrill cranial perforator with lot number 0611/21 was manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the perforator did not stop, perforating dura in 2 of the 4 burr holes.It was also reported that the device will not be returned for evaluation as it has been discarded by the customer.On follow-up, it was reported that the event happened during a craniotomy procedure for tumor that appears to be a meningioma.It was reported that the affected dura was removed as per procedure, there was no procedure delay or non-routine actions taken because of the event, and the patient had no further issues post-surgery.
|
| |
|
Search Alerts/Recalls
|
|
|