H3, h6: the navio drill, part number 101209, s/n (b)(6), intended for use in treatment was returned for evaluation.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The reported problem was visually confirmed.The exterior condition of the device showed scratches, dings, hose worn, fraying cable at connector cap.A functional evaluation was performed.The bur would not remain in place during the drill test.The most likely cause of this event is component failure.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
|