• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH NCB, FEMUR PLATE, RIGHT, 9 HOLES, 246 MM; NCB PLATING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH NCB, FEMUR PLATE, RIGHT, 9 HOLES, 246 MM; NCB PLATING SYSTEM Back to Search Results
Model Number N/A
Medical Device Problem Codes Patient Device Interaction Problem (4001); Migration (4003)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 08/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medical products: ncb, cancellous screw, 5.0 mm, 32 mm, 75 mm; catalog#: 02.03152.075; lot#: unknown.Ncb, screw, 5.0, 30 mm; catalog#: 02.03150.030; lot#: unknown.Ncb, cancellous screw, 5.0 mm, 32 mm, 65 mm; catalog#: 02.03152.065; lot#: unknown.Ncb, cancellous screw, 5.0 mm, 32 mm, 60 mm; catalog#: 02.03152.060; lot#: unknown.Ncb, locking cap; catalog#: 02.03150.300; lot#: unknown.Ncb, locking cap; catalog#: 02.03150.300; lot#: unknown.Ncb, locking cap; catalog#: 02.03150.300; lot#: unknown.Ncb, locking cap; catalog#: 02.03150.300; lot#: unknown.Ncb, locking cap; catalog#: 02.03150.300; lot#: unknown.Therapy date: (b)(6) 2021.The manufacturer received other source documents for review.The manufacturer did not receive the device for investigation.Where lot numbers were received for the devices, the device history records were reviewed and found to be conforming.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).
 
Event or Problem Description
Patient was implanted with an ncb distal femur plate on the right side and underwent a revision surgery since the screws had separated from the plate proximally(migrated).
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected and additional information.Additional: d9, h2, h6 correction: b4, g3, g6, h10.Event description: it was reported that the patient was implanted with ncb distal femur plate on (b)(6), 2020 and underwent a revision surgery on (b)(6) 2021 since the screws had separated from the plate proximally (migrated).Review of received data: - no medical data relevant to the case has been received.- due diligence: no further due diligence required as all required information to support the conclusion is available or was already requested.Product evaluation: - visual examination: the visual examination shows scratches on the surface of the plate.On three distal screw holes, there are signs of wear.One the 5 cancellous screws, there are no signs of damage.On the 3 self-tapping screws, there are signs of wear on each screw head.Review of product documentation: - device purpose: all involved devices are intended for treatment.- product compatibility: the product combination was approved by zimmer biomet.- dhr review: review of the device history records identified no deviations or anomalies during manufacturing.- ncr(s): no ncr with a potential correlation to the reported event was found.Conclusion: it was reported that the patient was implanted with ncb distal femur plate on (b)(6), 2020 and underwent a revision surgery on (b)(6) 2021 since the screws had separated from the plate proximally (migrated).Based on the investigation the reported event cannot be confirmed.The quality records show that all specified characteristics have met the specifications valid at the time of production.Therefore, the investigation did not identify a nonconformance or a complaint out of box (coob).Based on the given information and the results of the investigation, we were not able to identify a specific root cause for this issue.The need for corrective measures is not indicated and zimmer switzerland manufacturing gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
 
Event or Problem Description
No change to previously reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCB, FEMUR PLATE, RIGHT, 9 HOLES, 246 MM
Common Device Name
NCB PLATING SYSTEM
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12407267
Report Number0009613350-2021-00448
Device Sequence Number3661752
Product Code HRS
UDI-Device Identifier00889024297371
UDI-Public00889024297371
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K042695
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number02.03260.009
Device Lot Number2975382
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2021
Supplement Date Received by Manufacturer11/05/2021
Initial Report FDA Received Date09/01/2021
Supplement Report FDA Received Date11/12/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/28/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-