DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, SPANISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-58 |
Device Problem
Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device not accessible for testing: additional information is being requested in regards to whether or not the customer has requested for getinge to service the iabp unit involved.A supplemental report will be submitted when additional information is provided.
|
|
Event Description
|
It was reported that the cs300 intra-aortic balloon pump (iabp) had a blood pressure module failure and would not recognize the fiber optic sensor.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
|
Event Description
|
It was reported before patient use, that the cs300 intra-aortic balloon pump (iabp) had a blood pressure module failure and would not recognize the fiber optic sensor.There was no patient involvement.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate the issue.The fse replaced the lamps of the fiber optic module and then continued to perform functional tests in service mode.The equipment did not show any failures.However, based on the service, the equipment still presented failures in the control through the fiber optic.So, in order to solve the issue, the fse replaced the complete fiber optic assy.Then, the fse performed the fiber optic tests, which were satisfactory.The fse also conducted calibration tests and functional tests.The fse was then able confirmed that the unit was functional.
|
|
Event Description
|
N/a.
|
|