Model Number 0998-00-3023-58
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2021
Event Type  malfunction  
Manufacturer Narrative
Device not accessible for testing: additional information is being requested in regards to whether or not the customer has requested for getinge to service the iabp unit involved.A supplemental report will be submitted when additional information is provided.
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) had a blood pressure module failure and would not recognize the fiber optic sensor.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
 
Event Description
It was reported before patient use, that the cs300 intra-aortic balloon pump (iabp) had a blood pressure module failure and would not recognize the fiber optic sensor.There was no patient involvement.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate the issue.The fse replaced the lamps of the fiber optic module and then continued to perform functional tests in service mode.The equipment did not show any failures.However, based on the service, the equipment still presented failures in the control through the fiber optic.So, in order to solve the issue, the fse replaced the complete fiber optic assy.Then, the fse performed the fiber optic tests, which were satisfactory.The fse also conducted calibration tests and functional tests.The fse was then able confirmed that the unit was functional.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, SPANISH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12407495
MDR Text Key269601200
Report Number2249723-2021-01978
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108339
UDI-Public10607567108339
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-58
Device Catalogue Number0998-00-3023-58
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2011
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-