| Model Number |
87038 |
| Medical Device Problem Code |
Operating System Becomes Nonfunctional (2996)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
During preparation for a procedure it was reported that the device rhythmia mapping system overheated and crashed.The procedure was not completed due to this event.The event occurred during preparation.However, the sedation status of the patient was not specified.The patient is stable, no patient complication or other additional intervention reported.
|
| |
|
Event or Problem Description
|
|
During preparation for a procedure it was reported that the device arrhythmia mapping system overheated and crashed.The procedure was not completed due to this event.The event occurred during preparation.However, the sedation status of the patient was not specified.The patient is stable, no patient complication or other additional intervention reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The signal station was returned to boston scientific for analysis.Visual inspection did not reveal any abnormalities.During functional testing, the system overheated and a burning smell was observed.Error messages were also triggered on the work station.The reported observation was confirmed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|
| |
|