A director of quality/ risk management reported an issue with a bioflo midline 4f sl-20cm max sterile barrier nursing kit w df drape and two nit pg.During an insertion procedure, the patient's vein was accessed using the needle and the physician attempted to advance the guidewire but the patient moved and access was lost.The physician was unable to remove the guidewire; therefore, an interventional radiologist (ir) went to the patient's bedside to evaluate.The ir was able to remove the guidewire; however, the tip of the wire was not intact and an x-ray revealed that a piece was remaining in that patient's arm.The ir removed the wire and serial xrays of arm and chest were performed, as well as an ultrasound of the arm which revealed a thrombus.The patient was sent to (b)(6) medical center for further evaluation on (b)(6) 2021 with no further information at this time.The reported device is not available to be returned to the manufacturer for evaluation.
|
|
Event description updated with additional information: a director of quality/ risk management reported an issue with a bioflo midline 4f sl-20cm max sterile barrier nursing kit w df drape and two nit pg.The patient required a midline to be placed to treat sepsis, needing iv abx.During an insertion procedure, the patient's vein was accessed using the needle and the physician attempted to advance the guidewire but the patient moved and access was lost.The physician was unable to remove the guidewire; therefore, an interventional radiologist (ir) went to the patient's bedside to evaluate.The ir was able to remove the guidewire; however, the tip of the wire was not intact and an x-ray revealed that a piece was remaining in that patient's arm.The ir removed the wire tip and serial xrays of arm and chest were performed, as well as an ultrasound of the arm which revealed a thrombus.The procedure/treatment was aborted due to the thrombus, therefore the midline was not placed.The patient was then sent to (b)(6) medical center for further evaluation of the thrombus on (b)(6) 2021.Additional information reported that as of august 31, 2021 the patient was still at (b)(6) medical center.It is unknown if the thrombus is related to the device.
|
|
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.However, the customer provided an x-ray image of the reported defect.The guidewire accessory is supplied to angiodynamics by the supplier (b)(4).(b)(4) was notified of the event via scar004406 for a device history review of the supplier lot.As per the scar004406 response, "due to there not being pictures and samples available in the scar notification, a sample evaluation was not performed.A general training review was performed.It was verified that all personal was properly trained in all manufacturing process steps and in quality inspection as well".The customer's reported complaint description cannot be confirmed, no sample was returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.A potential root cause is end user pulling guidewire back through/against the needle.This can damage the guidewire tip against the sharp edge of the needle bevel and can result in tip detachment.This is cautioned against in the dfu.A review of the packaging device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: warning contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged.If damage is found, call your navilyst medical representative.Inspect prior to use to verify that no damage has occurred in shipping.For single patient use only.Do not reuse, reprocess or resterilize.Reuse, reprocessing or resterilization may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness or death.Operational instructions 1.Prior to insertion, flush all components with saline or heparinized/saline.2.Gain percutaneous access with the 21 g (0.9 mm) vascular introducer needle.3.Advance the 0.018 inch (0.46 mm) guidewire through the needle.Caution: the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.4.Withdraw the introducer needle, leaving the 0.018 inch (0.46 mm) guidewire in place.5.Advance the coaxial assembly over the 0.018 inch (0.46 mm) guidewire.6.Simultaneously remove the dilator and 0.018 inch (0.46 mm) guidewire, while holding the sheath portion of the assembly in place.7.Advance a 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) guidewire into the sheath.8.Remove the sheath, leaving the 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) wire.Precautions · do not advance or withdraw a guidewire against resistance until the cause of the resistance has been determined.Excessive force against resistance may result in guidewire damage or vessel perforation.Caution: the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|