COVIDIEN LLC STRATA II VALVE, REGULAR, WITH BIOGLIDE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 92866
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the procedure and before the valve implantation into the patient, the doctor checked the valve mechanically it's functionality and found that it was leaking on the delta chamber.The procedure concluded with another product.This resulted in a procedure delay of 20 minutes.The patient's status was alive-no injury.
 
Event or Problem Description
Additional information received reported that there maybe was a more harsh mechanical force to the valve by the physician during the valve check that lead to the leak.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
The returned valve was patent.The valve met the requirements for valve flux testing and reflux testing.A tear was observed on the delta chamber of the valve.The valve did not meet requirements of leakage testing due to the presence of this tear.It is unknown how this damage may have occurred.The ifu cautions, ¿low tear strength is a characteristic of most silicone elastomer materials.Improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting or crushing of components.Such damage may lead to loss of shunt integrity and necessitate premature surgical revision of the shunt system.¿ the valve did not meet requirements for siphon control and pressure flow/pre-implantation pressure testing.All valves are 100% tested at the time of manufacturing.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II VALVE, REGULAR, WITH BIOGLIDE
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
COVIDIEN LLC
15 hampshire street
mansfield 02048
Manufacturer (Section G)
COVIDIEN LLC
15 hampshire street
mansfield 02048
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key12407992
Report Number9612501-2021-01449
Device Sequence Number5292473
Product Code JXG
UDI-Device Identifier00763000027322
UDI-Public00763000027322
Combination Product (Y/N)N
Initial Reporter CountryGR
PMA/510(K) Number
K042465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/14/2024
Device Model Number92866
Device Catalogue Number92866
Device Lot Number0221442476
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2021
Supplement Date Received by Manufacturer09/09/2021
11/23/2021
Initial Report FDA Received Date09/01/2021
Supplement Report FDA Received Date09/15/2021
12/03/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-