• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TRIAL 20 DEG 36MM 58-60 G; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TRIAL 20 DEG 36MM 58-60 G; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71363658
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that two (2) trials, a trial 20 deg ant+4 36 58-60 g and a trial 20 deg 36mm 58-60 g are broken.It is unknown whether this problem was noticed in a surgical environment or not; therefore, patient involvement has not been confirmed.
 
Additional Manufacturer Narrative
H3, h6: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned trial 20 deg 36mm 58-60 g confirmed the device is fractured on the top.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.H3, h6, h10.
 
Additional Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable, damage may occur during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIAL 20 DEG 36MM 58-60 G
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12408125
Report Number1020279-2021-06667
Device Sequence Number3349928
Product Code JDG
UDI-Device Identifier03596010479839
UDI-Public03596010479839
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71363658
Device Catalogue Number71363658
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2021
Initial Date Received by Manufacturer 08/16/2021
Supplement Date Received by Manufacturer09/30/2021
11/01/2021
Initial Report FDA Received Date09/01/2021
Supplement Report FDA Received Date10/01/2021
11/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-