| Catalog Number |
07154003001 |
| Medical Device Problem Code |
Application Program Version or Upgrade Problem (2881)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter states there is an issue with their cobas infinity software.A rule was created in the cobas infinity software to block test results marked with a flag indicating that the sample is hemolytic.The reporter states that the rule is triggered sometimes and sometimes not.When the rule is not triggered, the software will not block the result and the result is sent to the customer's laboratory information system.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation determined there is an issue with low performance of the rule engine of the software.There is a delay in applying a rule to a result.This issue is causing the customer to think that the rule execution has failed, and therefore they have to manually perform the action that was supposed to be performed by this rule.If the result is not manually modified, the rule will eventually be executed and correctly modify the result.
|
| |
|