SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 5-6 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71453222
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 08/16/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a primary tka surgery had been performed on an unknown date, the patient experienced instability of the joint.A revision surgery was performed on (b)(6) 2021 to treat this adverse event; the lgn ps high flex xlpe sz 5-6 11mm was explanted and replaced with an unknown prosthesis.Surgeon stated that there is no fault related to the lgn ps high flex xlpe sz 5-6 11mm originally implanted.The patient outcome is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, smith and nephew has not received the device/ adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to fit/sizing issue, lifetime of device or patient condition.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 5-6 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12410961
Report Number1020279-2021-06688
Device Sequence Number3350549
Product Code JWH
UDI-Device Identifier00885556035351
UDI-Public00885556035351
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2001
Device Explanted Year2021
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2025
Device Model Number71453222
Device Catalogue Number71453222
Device Lot Number15LM03166
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/16/2021
Supplement Date Received by Manufacturer09/30/2021
Initial Report FDA Received Date09/01/2021
Supplement Report FDA Received Date10/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-