It was reported that, after a primary tka surgery had been performed on an unknown date, the patient experienced instability of the joint.A revision surgery was performed on (b)(6) 2021 to treat this adverse event; the lgn ps high flex xlpe sz 5-6 11mm was explanted and replaced with an unknown prosthesis.Surgeon stated that there is no fault related to the lgn ps high flex xlpe sz 5-6 11mm originally implanted.The patient outcome is unknown.
|
|
Internal complaint reference: (b)(4).The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, smith and nephew has not received the device/ adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to fit/sizing issue, lifetime of device or patient condition.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|