MAUDE Adverse Event Report: ENDOLOGIX AFX; VELA SUPRARENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX; VELA SUPRARENAL Back to Search Results
Model Number A28-28/C95-O20V
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.
 
Event or Problem Description
The patient was initially implanted with an afx2 bifurcated stent graft, two (2) suprarenal aortic extensions, and two (2) infrarenal aortic extensions to treat an abdominal aortic aneurysm (aaa).Approximately one (1) year post initial procedure there was a type 1a endoleak noted and a suspected type 2 (non-device-related) endoleak.The physician elected to coil embolize to resolve the type 1a endoleak and coil embolize the lumbar artery to resolve the type 2 endoleak.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.There are no other equivalent adverse events/incidents for this lot number existing within the endologix complaint handling system.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted in the patient.A clinical evaluation of the adverse event/incident could not be completed.No medical records nor medical imaging relevant to the reported adverse event/incident was received by endologix.The user facility was unable to provide this information.Due to the absence of medical records and medical imaging; device, use, procedure, and/or anatomy relatedness to this adverse event/incident could not be evaluated.The final patient status remains unknown.The final patient status was not made available to endologix.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with duraply g4: date of received by manufacturer - updated h6: result code: remove code 3233 h6: conclusion code: remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
VELA SUPRARENAL
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key12411494
Report Number2031527-2021-00326
Device Sequence Number3661491
Product Code MIH
UDI-Device Identifier00818009013637
UDI-Public(01)00818009013637(17)211108
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 08/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/08/2021
Device Model NumberA28-28/C95-O20V
Device Lot Number2009363006
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/17/2021
Supplement Date Received by Manufacturer09/10/2021
Initial Report FDA Received Date09/01/2021
Supplement Report FDA Received Date09/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BIFURCATED STENT GRAFT, LOT # 1973106009; INFRARENAL AORTIC EXTENSION, LOT # 1763600016; INFRARENAL AORTIC EXTENSION, LOT # 1812758006; SUPRARENAL AORTIC EXTENSION, LOT # 1953038010
Outcome Attributed to Adverse Event Required Intervention;
-
-