• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND SENTRANT INTRODUCER SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND SENTRANT INTRODUCER SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number SENSH2228W
Medical Device Problem Codes Difficult to Insert (1316); Leak/Splash (1354)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A sentrant sheath was used as an accessory device in a procedure.It was reported during the procedure, that the sheath was introduced into the patients left femoral artery over a 0.35 non mdt guide wire. while the sheath was being introduced, the physician  felt that there was too much leaking around the wire going through the dilator.Once the  sheath was introduced and dilator was removed, bleeding ceased.The physician felt that there was too much leaking in the dilator lumen around 0.35 wire.The patient was heparinized at the time of sheaths introduction into body.No cause was reported.  no additional clinical sequalae were provided and the patient is fine.
 
Additional Manufacturer Narrative
Additional information received: it was reported per the physician, there was no damage to dilator or sheath.It was suspected the inner lumen of dilator diameter exceeded the 035 wires ability to easily pass through and prevent bleeding through the dilator lumen.Per the physician; it usually is not an issue with sentrant insertion, however in this case, sheath insertion was difficult and time consuming which caused a fair amount of blood loss through the dilator.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTRANT INTRODUCER SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
EI 
MDR Report Key12412505
Report Number9612164-2021-03391
Device Sequence Number11697023
Product Code DYB
UDI-Device Identifier00643169792623
UDI-Public00643169792623
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K123990
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/24/2022
Device Model NumberSENSH2228W
Device Catalogue NumberSENSH2228W
Device Lot Number00151123
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/31/2021
Supplement Date Received by Manufacturer09/03/2021
Initial Report FDA Received Date09/02/2021
Supplement Report FDA Received Date09/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age89 YR
Patient Weight77
-
-