| Model Number |
45031 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that sterilty was compromised.An angiojet solent omni catheter was selected for use.During preparation, it was noted that the sterile package was already opened.The product was not sent to the sterile field.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that sterilty was compromised.An angiojet solent omni catheter was selected for use.During preparation, it was noted that the sterile package was already opened.The product was not sent to the sterile field.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the product return contained an angiojet solent omni.The device was returned inside of the packaging, with the white strip that indicates a sealed back missing.The assembly, supply line, shaft, saline supply and tip were all visually examined for potential damage.A functional test was performed.The pump was placed into the ultra drive console and the device primed with no complication.The device was run for a total time of 90 seconds in thrombectomy more.The device stayed within the expected range of the product, with an average pressure.No damage or discrepancies were noted on the rest of the catheter upon visual inspection.
|
| |
|