MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ANGIOJET SOLENT OMNI; CATHETER, EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ANGIOJET SOLENT OMNI; CATHETER, EMBOLECTOMY Back to Search Results
Model Number 45031
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that sterilty was compromised.An angiojet solent omni catheter was selected for use.During preparation, it was noted that the sterile package was already opened.The product was not sent to the sterile field.The procedure was completed with another of the same device.No patient complications were reported.
 
Event or Problem Description
It was reported that sterilty was compromised.An angiojet solent omni catheter was selected for use.During preparation, it was noted that the sterile package was already opened.The product was not sent to the sterile field.The procedure was completed with another of the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the product return contained an angiojet solent omni.The device was returned inside of the packaging, with the white strip that indicates a sealed back missing.The assembly, supply line, shaft, saline supply and tip were all visually examined for potential damage.A functional test was performed.The pump was placed into the ultra drive console and the device primed with no complication.The device was run for a total time of 90 seconds in thrombectomy more.The device stayed within the expected range of the product, with an average pressure.No damage or discrepancies were noted on the rest of the catheter upon visual inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET SOLENT OMNI
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12413648
Report Number2134265-2021-11113
Device Sequence Number3577259
Product Code KRA
UDI-Device Identifier08714729889663
UDI-Public08714729889663
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K111182
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/21/2023
Device Model Number45031
Device Catalogue Number45031
Device Lot Number0027231493
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2021
Initial Date Received by Manufacturer 08/16/2021
Supplement Date Received by Manufacturer09/08/2021
Initial Report FDA Received Date09/02/2021
Supplement Report FDA Received Date09/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-