Model Number 10094139
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the artis zee mp system.During an interventional procedure, the user reported that the system went into bypass mode.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The investigation showed that the image acquisition system (ias) failed due to a defect in the ias computer, which was confirmed by the evaluation of the log files.As a result, the system switched to the bypass fluoro mode, as intended for such a case, and displayed a corresponding error message to the operator.This special mode, which is a mitigation of the problem, allows the system to continue generating imaging.However, neither acquisition nor storage or further processing of data is possible, and the image quality is reduced.Such an error can only be eliminated by replacing the affected hardware.The entire ias bb3.5 pc was replaced during service intervention.The system is running again as specified.The spare parts consumption of the affected part was checked.A possible error accumulation or even a possible general error, which would require corrective measures of the installed base, could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE MP
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH- AT
siemensstrasse 1
forchheim, germany 91301
GM  91301
MDR Report Key12413670
Report Number3004977335-2021-94259
Device Sequence Number17059233
Product Code OWB
UDI-Device Identifier04056869010076
UDI-Public04056869010076
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K181407
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial,Followup
Report Date (Section B) 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10094139
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/23/2021
Supplement Date Received by Manufacturer10/15/2021
Initial Report FDA Received Date09/02/2021
Supplement Report FDA Received Date10/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-