| Model Number |
490100 |
| Medical Device Problem Code |
Contamination of Device Ingredient or Reagent (2901)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medical device expiration date: na.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd prepstain¿ fungal contamination has been occurring.There was no report of patient impact.The following information was provided by the initial reporter: it was reported that the customer reports that they have been experiencing fungal contamination but cannot determine which prepstain instrument is responsible.
|
| |
|
Additional Manufacturer Narrative
|
|
After further evaluation of the complaint, it has been determined that the previously submitted mfr report# was sent in error.Fungal contamination for women's health prepstain is not considered to be a reportable malfunction.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd prepstain¿ fungal contamination has been occurring.There was no report of patient impact.The following information was provided by the initial reporter: it was reported that the customer reports that they have been experiencing fungal contamination but cannot determine which prepstain instrument is responsible.
|
| |
|