• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNIVERSAL FEM SLV DIS POR 40MM; UNIVERSAL KNEE STEMS AND SLEEVES : KNEE FEMORAL ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNIVERSAL FEM SLV DIS POR 40MM; UNIVERSAL KNEE STEMS AND SLEEVES : KNEE FEMORAL ACCESSORY Back to Search Results
Model Number 1294-53-235
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Pain (1994); Limb Fracture (4518); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 04/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical adverse event received for bone fracture of medial and lateral femoral condlyles: abnormal radiographic evaluation - likely femoral component loosening (unconfirmed), with fracture of the medial and lateral condyles, and instability event is serious and is considered severe event is definitely related to device and is definitely not related to procedure.Date of implantation: (b)(6) 2014.Date of revision: (b)(6) 2017 (all components).Date of event (onset): (b)(6) 2021.(left knee).Treatment: revision of the femoral and insert components occurred on (b)(6) 2021.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL FEM SLV DIS POR 40MM
Common Device Name
UNIVERSAL KNEE STEMS AND SLEEVES : KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12415381
Report Number1818910-2021-19231
Device Sequence Number17013426
Product Code JWH
UDI-Device Identifier10603295026327
UDI-Public10603295026327
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1294-53-235
Device Catalogue Number129453235
Device Lot Number515890
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/19/2021
Supplement Date Received by Manufacturer09/20/2021
Initial Report FDA Received Date09/02/2021
Supplement Report FDA Received Date09/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEPUY CMW 3 40G; DEPUY CMW 3 40G; MBT REVISION CEM TIB TRAY SZ 5; MBT TRAY SLEEVE POR M/L 61MM; PFC SIGMA FEM POST AUG SZ4 4MM; PFC SIGMA FEM POST AUG SZ4 8MM; PFC*SIGMA TC3 FEM LEFT SZ4; PFC*SIGMA/OV/DOME PAT 3PEG,41; SIG FEM ADAP +2/-2 OFFSET BOLT; SIGMA FEM ADAPTER 5 DEGREE; TC3 RP TIBIAL INSERT S4,25.0; UNIVERSAL FEM SLV DIS POR 40MM; DEPUY CMW 3 40G; DEPUY CMW 3 40G; MBT REVISION CEM TIB TRAY SZ 5; MBT TRAY SLEEVE POR M/L 61MM; PFC SIGMA FEM POST AUG SZ4 4MM; PFC SIGMA FEM POST AUG SZ4 8MM; PFC*SIGMA TC3 FEM LEFT SZ4; PFC*SIGMA/OV/DOME PAT 3PEG,41; SIG FEM ADAP +2/-2 OFFSET BOLT; SIGMA FEM ADAPTER 5 DEGREE; TC3 RP TIBIAL INSERT S4,25.0; UNIVERSAL FEM SLV DIS POR 40MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient Weight100
-
-