| Model Number |
96-0886 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Fall (1848); Pain (1994); Limb Fracture (4518); Joint Laxity (4526); Swelling/ Edema (4577)
|
| Date of Event |
04/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Clinical adverse event received for bone fracture of medial and lateral femoral condlyles: abnormal radiographic evaluation - likely femoral component loosening (unconfirmed), with fracture of the medial and lateral condyles, and instability event is serious and is considered severe event is definitely related to device and is definitely not related to procedure.Date of implantation: (b)(6) 2014.Date of revision: 03/31/2017 (all components).Date of event (onset): (b)(6) 2021.(left knee).Treatment: revision of the femoral and insert components occurred on 17/aug/2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|