| Model Number |
7580 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned product consisted of a maverick 2mr balloon catheter.The device was microscopically and visually examined.There was both contrast and blood in the inflation lumen and balloon.There was blood in the guidewire lumen and the balloon was loosely folded.The device was soaked for a period of time to help break up the contrast and blood and was then prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There was a pinhole in the balloon located 5mm from the tip of the device near the distal markerband.There was no visible markerband damage to the device.The device failed to inflate to rated burst pressure.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2021.It was reported that the device was damaged.The 70% stenosed target lesion was located in the mid left anterior descending artery.A 2.00mm x 12mm maverick balloon catheter was selected for use.However, during unpacking, the device was noted to be damaged.The procedure was completed with another of same device.No patient complications were reported and the patient was stable.However, returned device analysis revealed the device had been inside the patient and had a balloon pinhole.
|
| |
|