|
|
| Model Number |
SPM-35-W |
| Medical Device Problem Codes |
Degraded (1153); Material Deformation (2976); Mechanics Altered (2984); Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Abscess (1690); Adhesion(s) (1695); Bacterial Infection (1735); Emotional Changes (1831); Foreign Body Reaction (1868); Unspecified Infection (1930); Inflammation (1932); Nerve Damage (1979); Pain (1994); Hernia (2240); Impaired Healing (2378); Fluid Discharge (2686); Unspecified Tissue Injury (4559)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a right inguinal hernia.It was reported that after implant, the patient experienced nerve injury and pain.Post-operative patient treatment included pain meds for three weeks.
|
| |
|
Event or Problem Description
|
|
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a right inguinal hernia.It was reported that after implant, the patient experienced inflammation, bacterial proliferation, infection, mesh degradation, mesh encased in biofilm, mesh deformed, mesh not incorporated, chronic draining abscess, defective device, recurrence, mental and physical pain, disability, impairment, loss of enjoyment of life, nerve injury, and pain.Post-operative patient treatment included pain medications for three weeks.It was noted that the mesh was unable to be removed due to high risk.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional info - b5, b7, g1, h6 (patient codes) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a right inguinal hernia.It was reported that after implant, the patient experienced chronic pain, adhesions, inflammation, bacterial proliferation, infection, mesh degradation, mesh encased in biofilm, mesh deformed, mesh not incorporated, chronic draining abscess, defective device, recurrence, mental and physical pain, disability, impairment, loss of enjoyment of life, nerve injury, and pain.Post-operative patient treatment included pain medications for three weeks.It was noted that the mesh was unable to be removed due to high risk of losing testicle.
|
| |
|
Search Alerts/Recalls
|
|
|