• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL+ UD 8X18IN 4-0 S/A RB-1 CR; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ UD 8X18IN 4-0 S/A RB-1 CR; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number VCP714D
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Did the needle break at the tip or at the body? procedure date? a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 275 g/m.
 
Event or Problem Description
It was reported that a patient underwent a craniotomy procedure on (b)(6) 2021 and suture was used.During the procedure, the needle broke while sewing external closure of skull.Another like device was used to complete the procedure.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 11/03/2021.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigation summary: needle analysis: a failed suture needle, labeled as (b)(4), was submitted for fractographic evaluation.The component was identified as product code vcp714d.It was requested to assess the fracture mode of the failure.A fracture was observed at the tip of the needle.One side of the needle was received, the mating fracture surface was not provided for this evaluation.A scanning electron microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation of (b)(4) revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage observed coincidental to the fracture provides additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the tip of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.In order to avoid this kind of damage grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.The following information was requested, but unavailable: did the needle break at the tip or at the body? unk.Procedure date? unk.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 8X18IN 4-0 S/A RB-1 CR
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
MDR Report Key12418016
Report Number2210968-2021-07988
Device Sequence Number3476958
Product Code GAM
UDI-Device Identifier10705031055018
UDI-Public10705031055018
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberVCP714D
Device Catalogue NumberVCP714D
Device Lot NumberRAMBKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2021
Initial Date Received by Manufacturer 08/12/2021
Supplement Date Received by Manufacturer10/08/2021
Initial Report FDA Received Date09/02/2021
Supplement Report FDA Received Date11/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-