| Catalog Number |
0250070600 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the device clogged.
|
| |
|
Event or Problem Description
|
|
It was reported that the device clogged.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: today i had a customer that is very familiar with our ahto irrigator go through 4 consecutive ahto tubesets because of clogging in the body of the handpiece during suction.The issue was resolved when a different lot number was pulled.The probable root cause could bio burden substance from the use of the instrument port, or normal use.
|
| |
|