| Model Number |
V1390 |
| Medical Device Problem Code |
Free or Unrestricted Flow (2945)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: rn reports that "the tubing is stiffer and stickier, making it cumbersome to handle at times.The roller clamp is now made of a harder/sharper plastic and the roller is more difficult to roll up and down, which makes it both painful and inaccurate when calculating the drip rate.The roller clamp has literally sent shocking pain up my thumb, which has never happened with previous clamps, and it is either all the way open or extremely slow in many cases." no injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).Two (2) photographs were provided for evaluation.Based on the data from the investigation, the difference in product appearance was the result of an approved product material change and testing confirmed the device met all required performance specifications with no deviations or non-conformances identified during testing or manufacturing.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|