• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EVOS 2.5MM X 3.5MM DRILL GUIDE; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EVOS 2.5MM X 3.5MM DRILL GUIDE; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Model Number 71175041
Medical Device Problem Codes Break (1069); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during internal fixation surgery, the drill bit got stuck in the evos 2.5mm x 3.5mm drill guide and broke.Surgery was resumed, without any delay, with a back-up device.The patient was not injured as a consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device could not confirm the stated failure.There is no drill bit stuck inside.The visual did confirm burrs and gouges on both the 2.5mm x 3.5mm sides that would cause a drill bit to get stuck inside the device.The device shows signs of significant wear and use.A review of complaint history did not reveal additional complaints for the listed batch for the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS 2.5MM X 3.5MM DRILL GUIDE
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12421637
Report Number1020279-2021-06721
Device Sequence Number3662055
Product Code FZX
UDI-Device Identifier00885556636404
UDI-Public00885556636404
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71175041
Device Catalogue Number71175041
Device Lot Number17MSC0020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2021
Initial Date Received by Manufacturer 08/17/2021
Supplement Date Received by Manufacturer09/28/2021
Initial Report FDA Received Date09/03/2021
Supplement Report FDA Received Date09/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-