| Model Number |
295025-001 |
| Medical Device Problem Codes |
Battery Problem (2885); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The freedom onboard battery will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer, a syncardia authorized distributor, reported that patient's freedom onboard battery discharged immediately as soon as it was inserted into the freedom driver after a complete charge.
|
| |
|
Additional Manufacturer Narrative
|
|
The freedom onboard battery was evaluated per the freedom onboard battery evaluation procedure and passed all sections.During investigation testing, the customer-reported issue was not reproduced and there was no evidence of a device malfunction.The freedom driver as well as the freedom battery charger used at the time of the reported event were not returned and therefore could not be included in this investigation.The root cause of the customer-reported issue that the freedom onboard battery discharged immediately after it was inserted into the freedom driver after a complete charge could not be conclusively determined.The battery performed as intended.This issue will be tracked and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(6).(b)(4)follow-up report 1.
|
| |
|