• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ARTHROSCOPE Back to Search Results
Model Number 375-545-000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned to stryker sustainability solutions for evaluation.As the device was not returned for evaluation, inspection was unable to be performed, however, the complainant did photograph the reported failure of the shaver.Per the photograph provided by the complainant the reported event was confirmed.A review of the dhr for the reported lot number supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to user applied too much force to the device or the arthroscopic shaver may have come into contact with staples, clips or another metal object, resulting in damage to the blade.The instructions for use (ifu) state: before beginning the procedure, verify compatibility of all instruments and accessories.Plug in and set up the generator according to the instructions in the manufacturers¿ manual.Select an arthroscopic shaver with size, blade, and function most appropriate for the procedure.Inspect the instrument for overall condition and physical integrity.Do not use the instrument if any damage is noted.Return the instrument and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.Follow a suitable surgery protocol.Careful handling of the instrument is necessary to avoid damage or breakage as a result of excessive force.Do not apply excessive pressure or ¿side-load¿ the blade during use.Side-loading does not improve the performance of the instrument, can dull the blade, and/or produce metal particulates.Do not allow the arthroscopic shaver to come into contact with staples, clips or any metal object to avoid damage to the blade and possible patient injury.The tip of the bur or cutter must be irrigated periodically (general recommendation: once a minute) to cool the blade and prevent excised tissues from accumulating.Do not run the instrument without appropriate suction for the duration of the process.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
 
Event or Problem Description
It was reported the shaver tip broke off during the arthroscopy shoulder subacromial decompression case.The tip fell in the patient and was retrieved with an arthroscopic grasper.It was reported it took 10 seconds to retrieve the piece.The device was used for 5 minutes prior to breakage.There was no patient injury or medical intervention and extended procedure time reported was 2 minutes.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
MDR Report Key12422436
Report Number0001056128-2021-00050
Device Sequence Number14279952
Product Code HRX
UDI-Device Identifier00885825014223
UDI-Public00885825014223
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K012635
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/22/2024
Device Model Number375-545-000
Device Catalogue Number375-545-000RR
Device Lot Number12426488
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/06/2021
Initial Report FDA Received Date09/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-