• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORCYM CANADA CORP. MITROFLOW DLA AORTIC PERICARDIAL HEART VALVE; TISSUE HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORCYM CANADA CORP. MITROFLOW DLA AORTIC PERICARDIAL HEART VALVE; TISSUE HEART VALVE Back to Search Results
Model Number DLA23
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The manufacturing and material records for the mitroflow bioprosthetic pericardial heart valve, model #: dla23, s/n #:(b)(4), as they pertain to the reported event, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this valve satisfied all material, visual, and performance standards required for a dla23 mitroflow aortic pericardial heart valve at the time of manufacture and release.Since the device remains implanted, no further investigation is possible at this time.As reported in the scientific literature, structural valve deterioration is the major cause of failure of bioprosthetic heart valves.It is possible that the patient¿s clinical history and risk factors may have contributed to the structural valve deterioration observed in this mitroflow dl valve.However, since no further information is available, a definitive root cause cannot be stated at this time.It should be noted that structural valve deterioration is listed as a possible adverse event in the mitroflow ifu.The event is, therefore, a known inherent risk of the device.The manufacturer is following up to retrieve additional information on the event reported.Should the manufacturer receive further information, an update to this reporting activity will be provided.
 
Event or Problem Description
On (b)(6) 2021, an event of svd on a mitroflow dla23 valve was reported.The implant time and patient impact are presently unknown.
 
Additional Manufacturer Narrative
Corrected sections: h3, h6 (included codes: device not returned-4114; and no findings available-3221).Since the device was not returned to the manufacturer, no further investigation is possible at this time.Based on the limited information available, it is not possible to draw a definitive conclusion to the reported event.However, from the document review performed, no manufacturing deficiencies were identified.The manufacturer has contacted the physician for further information, but none was received to date.Given that the device functionality over time, patient's clinical history and risk factors, treatment and patient's impact are unknown, the root cause remains unknown at this time.It should be noted that structural valve deterioration is listed as a possible adverse event in the mitroflow valve ifu.The event is, therefore, a known inherent risk of the device.Should any additional information be received in the future, the manufacturer will provide an update to this reporting activity.H3 other text : unknown device disposition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITROFLOW DLA AORTIC PERICARDIAL HEART VALVE
Common Device Name
TISSUE HEART VALVE
Manufacturer (Section D)
CORCYM CANADA CORP.
5005 north fraser way
burnaby, bc
MDR Report Key12422449
Report Number3004478276-2021-00237
Device Sequence Number11590317
Product Code LWR
UDI-Device Identifier00896208000177
UDI-Public(01)00896208000177(240)DLA23(17)170531
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P060038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2017
Device Model NumberDLA23
Device Catalogue NumberDLA23
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/05/2021
Supplement Date Received by Manufacturer10/19/2021
Initial Report FDA Received Date09/03/2021
Supplement Report FDA Received Date10/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-