MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS VITAMIN D TOTAL II; VITAMIN D TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS VITAMIN D TOTAL II; VITAMIN D TEST SYSTEM Back to Search Results
Catalog Number 07028148190
Medical Device Problem Codes High Test Results (2457); Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Na.
 
Event or Problem Description
The initial reporter stated they received discrepant results for two patient samples tested with the elecsys vitamin d total gen.2 assay on a cobas 8000 e 602 module.It was asked, but it is not known if any incorrect results were reported outside of the laboratory.No units of measure were provided.The samples were initially tested using the vitamin d gen.2 assay and were frozen after measurement.The samples were then thawed to use as part of a validation study for the vitamin d gen.3 assay and were repeated using both the gen.2 and gen.3 assays.The measurements performed with the frozen/thawed sample were lower compared to the initial measurements.The vitamin d gen.3 assay or similar device is not cleared or approved for use in the united states.The first sample initially resulted in a vitamin d gen.2 value of 128 and then the sample was frozen.The sample was thawed and repeated using the gen.2 assay, resulting in a values of 91.6.The thawed sample was repeated using the gen.2 assay on a second analyzer, resulting as 78.6.The thawed sample was repeated using the gen.3 assay, resulting in a value of 83.9.The second sample initially resulted in a vitamin d gen.2 value of 157 and then the sample was frozen.The sample thawed and repeated using the gen.2 assay on a second analyzer, resulting in a value of 95.1.The thawed sample was repeated using the gen.3 assay, resulting in a value of 91.2.The e 801 analyzer serial number is (b)(4).
 
Additional Manufacturer Narrative
The units of measure used for the vitamin d assays are ng/ml.
 
Additional Manufacturer Narrative
The first sample was initially tested on (b)(6) 2021 and the second sample was initially tested on (b)(6) 2021.The customer confirmed they did not freeze the samples in question.The samples are stored on one module of the pre-analytics system for 2 hours after measurement and then they go to a second module for storage at 2-8 degrees celsius.Storage occurred overnight at 2-8 degrees celsius.Samples were repeated the following day.Calibration signals from the complained analyzer were within expected values.Calibration values from the second analyzer (e 801 serial number (b)(6)) were lower than expected.Quality controls were within range on the dates of the event.A general reagent issue could be excluded.Upon review of the alarm trace, no relevant alarms were observed.The investigation could not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS VITAMIN D TOTAL II
Common Device Name
VITAMIN D TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key12422624
Report Number1823260-2021-02591
Device Sequence Number19727141
Product Code MRG
UDI-Device Identifier04015630940127
UDI-Public04015630940127
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K162840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07028148190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2021
Supplement Date Received by Manufacturer09/20/2021
11/22/2021
Initial Report FDA Received Date09/03/2021
Supplement Report FDA Received Date09/20/2021
12/13/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-