• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; PLATE Back to Search Results
Model Number 50-795-26-09
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Impaired Healing (2378)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a plate broke.It was removed and replaced.
 
Additional Manufacturer Narrative
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number was not identified; therefore, the device history records were not reviewed.The failure root cause cannot be determined due to the device not being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
PLATE
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM  78570
Manufacturer (Section G)
KLS MARTIN L.P.
p.o. box 16369
jacksonville FL 32245
Manufacturer Contact
jennifer damato
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key12422757
Report Number9610905-2021-00078
Device Sequence Number17038210
Product Code MQN
UDI-Device Identifier00888118078896
UDI-Public(01)00888118078896
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K032442
Exemption Number2017029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number50-795-26-09
Device Lot Number233616
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2021
Supplement Date Received by Manufacturer11/19/2021
Initial Report FDA Received Date09/03/2021
Supplement Report FDA Received Date11/24/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age70 YR
Patient SexFemale
-
-