ABBOTT MEDICAL PROCLAIM 7 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG Back to Search Results
Model Number 3662
Device Problem Wireless Communication Problem (3283)
Patient Problem Inadequate Pain Relief (2388)
Event Date 08/18/2021
Event Type  Injury  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
 
Event Description
It was reported after a surgery the ipg cannot communicate with external devices.Surgical intervention may be pending.
 
Manufacturer Narrative
An inoperable implant was reported to abbott.The system was not set to surgery mode while the patient underwent an unrelated surgery where electro-cautery may have been used.The implant was not returned for analysis.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 7 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
MDR Report Key12423777
MDR Text Key269882696
Report Number1627487-2021-16813
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020222
UDI-Public05415067020222
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/21/2021
Device Model Number3662
Device Catalogue Number3662
Device Lot Number6814296
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/18/2021
Initial Date FDA Received09/03/2021
Supplement Dates Manufacturer Received09/20/2021
Supplement Dates FDA Received09/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number1627487/06/02/2017/001-C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age48 YR
-
-