| Model Number |
DLA23 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The manufacturing and material records for the mitroflow bioprosthetic pericardial heart valve, model # dla23, s/n # (b)(4), as they pertain to the reported event, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this valve satisfied all material, visual, and performance standards required for a dla23 mitroflow aortic pericardial heart valve at the time of manufacture and release.Since the device remains implanted, no further investigation is possible at this time.As reported in the scientific literature, structural valve deterioration is the major cause of failure of bioprosthetic heart valves.It is possible that the patient¿s clinical history and risk factors may have contributed to the structural valve deterioration observed in this mitroflow dl valve.However, since no further information is available, a definitive root cause cannot be stated at this time.It should be noted that structural valve deterioration is listed as a possible adverse event in the mitroflow ifu.The event is, therefore, a known inherent risk of the device.The manufacturer is following up to retrieve additional information on the event reported.Should the manufacturer receive further information, an update to this reporting activity will be provided.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, an event of svd on a mitroflow dla23 valve was reported.The implant time and patient impact are presently unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected sections: h6 (to include codes 4114-device not returned and 3221-no findings available); since the device was not returned to the manufacturer, no further investigation is possible at this time.Based on the limited information available, it is not possible to draw a definitive conclusion to the reported event.However, from the document review performed, no manufacturing deficiencies were identified.The manufacturer has contacted the physician for further information, but none was received to date.Given that the device functionality over time, patient's clinical history and risk factors, treatment and patient's impact are unknown, the root cause remains unknown at this time.It should be noted that structural valve deterioration is listed as a possible adverse event in the mitroflow valve ifu.The event is, therefore, a known inherent risk of the device.Should any additional information be received in the future, the manufacturer will provide an update to this reporting activity.
|
| |
|
Search Alerts/Recalls
|
|