The manufacturing and material records for the mitroflow bioprosthetic pericardial heart valve, model # lxa25, s/n # (b)(4), as they pertain to the reported event, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this valve satisfied all material, visual, and performance standards required for a lxa25 mitroflow aortic pericardial heart valve at the time of manufacture and release.Since the device was not returned to the manufacturer, no further investigation is possible at this time.As reported in the scientific literature, structural valve deterioration is the major cause of failure of bioprosthetic heart valves.It is possible that the patient¿s clinical history and risk factors may have contributed to the structural valve deterioration observed in this mitroflow lx valve.However, since no further information is available, a definitive root cause cannot be stated at this time.It should be noted that structural valve deterioration is listed as a possible adverse event in the mitroflow ifu.The event is, therefore, a known inherent risk of the device.The manufacturer is following up to retrieve additional information on the event reported.Should the manufacturer receive further information, an update to this reporting activity will be provided.
|