| Model Number |
352506100E |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The sample is indicated as available for evaluation.A follow-up report will be submitted once the sample has been received and reviewed.
|
| |
|
Event or Problem Description
|
|
Distal radiopaque band came off the dilator while getting inserted into the body, and the radiopaque band traveled into the right lungs¿.
|
| |
|
Event or Problem Description
|
|
Distal radiopaque band came off the dilator while getting inserted into the body, and the radiopaque band traveled into the right lungs¿ ¿ see photos.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned.Photos provided by the customer showed the distal side of the dilator marker missing, and the complaint was confirmed.Capa (c-2021-035) shall address a customer complaint regarding the option elite dilator's platinum marker band detaching during a procedure.The capa will investigate why the platinum marker did not stay secured in position and implement a solution to resolve the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the dhr and inspection records was conducted, and no similar concerns were found.According to the photo, provided by the customer, found one of two marker, distal side of the dilator marker is missing, and the complaint was confirmed.Capa (c-2021-035) shall address a customer complaint regarding the option elite dilator's platinum marker band detaching during a procedure.The capa will investigate why the platinum marker did not stay secured in position and implement a solution to resolve the issue.
|
| |
|
Event or Problem Description
|
|
Distal radiopaque band came off the dilator while getting inserted into the body, and the radiopaque band traveled into the right lungs¿ ¿ see photos.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the dhr and inspection records was conducted, and no similar concerns were found.Quality engineer and complaint analyst reviewed the sample returned from the customer.Customer returned dilator and found one of two marker band was missing (distal side of the dilator) and the complaint was confirmed.Capa (c-2021-035) shall address a customer complaint regarding the option elite dilator's platinum marker band detaching during a procedure.The capa will investigate why the platinum marker did not stay secured in position and implement a solution to resolve the issue.
|
| |
|
Event or Problem Description
|
|
Distal radiopaque band came off the dilator while getting inserted into the body, and the radiopaque band traveled into the right lungs¿ ¿ see photos.
|
| |
|