| Model Number |
SVMB13 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Abscess (1690); Unspecified Infection (1930); Pain (1994)
|
| Date of Event |
06/28/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Multiple mdr reports were filed for this event.Please see associated reports: 0002023141-2021-02457 and 0002023141-2021-02458.Zimmer biomet complaint number (b)(4).Weight unknown / not provided.Initial reporter fax number unknown / not provided.Additional pma/510(k) number ¿ k011028|k013227.A summary investigation has been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Should additional information be received which indicates that the device may have caused or contributed to the infection event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
Doctor indicated infection.Patient presented with infection, pain and abscess.Patient not returning for future placement.Tooth sites #21-22-11.
|
| |
|
Event or Problem Description
|
|
No additional event information at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Multiple mdr reports were filed for this event.Please see associated reports: 0002023141-2021-02457-1 and 0002023141-2021-02458-1.This report is being submitted to relay additional information and device evaluation.The following sections are being reported: h6: type of investigation code was added: 3331.Dhr review was completed for the subject lot number (1240770 & 63738353).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Sterilization record (op160 & st3285) was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number (1240770 & 63738353) revealed that there are no existing non-conformances/capa/hhe/d/ie/product holds for the reported product for similar events (complaint category keyword: infection).October post market trending was reviewed and there were no actionable events or corrective actions for the reported event (infection) or for the reported device (tsvb11 x 2 & svmb13).As documented in the summary investigation, contributing factors for the reported event likely exist outside of zimmer biomet control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigation, no malfunction occurred upon investigation.The reported event remains non-verifiable as it is a medical condition.
|
| |
|