| Model Number |
24674 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Initial reporter facility name: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The stenosed target lesion was located in a shunt.A 6.0 x 80, 135cm mustang balloon catheter was advanced for dilatation.However, during inflation at 18 atmospheres, the balloon ruptured.Another 6.0 x 80, 135cm mustang balloon catheter was advanced but it also ruptured during inflation at 18 atmospheres.The procedure was completed with a different device.No patient complications were reported and the patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
(a2) age at time of event: 18 years or older.(e1) initial reporter facility name: (b)(6).Updated b5: describe event or problem.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The stenosed target lesion was located in a shunt.A 6.0 x 80, 135cm mustang balloon catheter was advanced for dilatation.However, during inflation at 18 atmospheres, the balloon ruptured.Another 6.0 x 80, 135cm mustang balloon catheter was advanced but it also ruptured during inflation at 18 atmospheres.The procedure was completed with a different device.No patient complications were reported and the patient status was stable.It was further reported that the target lesion was about 30% stenosed and was located in the superficial femoral artery.Both devices were not inflated to the nominal bar at about 5 seconds before it ruptured.Both devices were completely removed and was removed directly from the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The stenosed target lesion was located in a shunt.A 6.0 x 80, 135cm mustang balloon catheter was advanced for dilatation.However, during inflation at 18 atmospheres, the balloon ruptured.Another 6.0 x 80, 135cm mustang balloon catheter was advanced but it also ruptured during inflation at 18 atmospheres.The procedure was completed with a different device.No patient complications were reported and the patient status was stable.It was further reported that the target lesion was about 30% stenosed and was located in the superficial femoral artery.Both devices were not inflated to the nominal bar at about 5 seconds before it ruptured.Both devices were completely removed and was removed directly from the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(a2) age at time of event: 18 years or older.(e1) initial reporter facility name: (b)(6) hospital.Device evaluated by mfr.: a mustang was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A longitudinal tear was identified in the balloon material.The tear measured 43mm in length and extended from a position 6mm proximal of the proximal markerband to 39mm proximal of the distal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both markerbands were undamaged and present on the shaft of the device.No other issues were identified during the product analysis.
|
| |
|