It was reported that on (b)(6) 2021 during dialysis, venous pressure increased and when contrast imaging was performed the area from the loop became invisible.A percutaneous transluminal angioplasty (pta) was performed the next day.The graft was occluded again a week later, and the occluded area was to be replaced.However, when the vein was incised and the graft was excised, the outer and inner layers were separated.The entire graft was therefore replaced.
|
|
Manufacturer's investigation conclusion: the report of occlusion and peeling of the graft layers could not be confirmed.The customer reported that the graft was originally implanted on (b)(6) 2016.During dialysis on (b)(6) 2021, venous pressure increased and the area of the loop was not visible when contrast imaging was performed.The issue was addressed via percutaneous transluminal angioplasty but the graft became occluded again one week later.During surgery to replace the occluded area, the outer and inner layers of the graft reportedly became separated.The entire graft was subsequently replaced.No product was returned for investigation.The cause of the reported occlusion could not be determined.The vectra vag instructions for use cautions that when using a balloon angioplasty or embolectomy catheter within the lumen of the vectra vag, match the inflated balloon size to the inner diameter size of the vag.Over-inflation of the balloon or use of an inappropriately sized balloon catheter may damage the graft.Care must be exercised to avoid causing excessive axial elongation of the graft during retraction.The thrombectomy section of the ifu also states to follow the catheter manufacturer¿s instructions regarding size, selection, and balloon inflation, matching the balloon size to the inner diameter of the graft.Over-inflation and excessive pulling may dilate or damage the graft.Graft preparation and implantation are addressed in the vectra vascular access graft (vag) instructions for use (ifu).The precautions section states that when using a balloon angioplasty or embolectomy catheter within the lumen of the vectra vag, match the inflated balloon size to the inner diameter size of the vag.Over-inflation of the balloon or use of an inappropriately sized balloon catheter may damage the graft.Care must be exercised to avoid causing excessive axial elongation of the graft during retraction.The thrombectomy section states to follow the catheter manufacturer¿s instructions regarding size, selection, and balloon inflation, matching the balloon size to the inner diameter of the graft.Over-inflation and excessive pulling may dilate or damage the graft.If devices such as adherent clot catheter are used to declot the vectra grafts, it is important to match the catheter size to the internal diameter of the graft.If a longitudinal incision is used, place stay stitches at each end of the incision before introducing the embolectomy catheter.If a transversive incision is used, no stay stitch is necessary and a horizontal mattress suture technique will aid closure.Occlusion, stenosis, and thrombosis are listed as potential complications with the use of a vascular prosthesis.Intragraft obstruction is listed as a potential complication in table 2 in the summary of vectra vag clinical experience section of the ifu.Review of the device history records showed no deviations from manufacturing or qa specifications, including visual and dimensional inspection.No further information was provided.The manufacturer is closing the file on this event.
|