ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT VASCULAR STENT Back to Search Results
Model Number EX071503CS
Medical Device Problem Codes Entrapment of Device (1212); Fracture (1260); Misfire (2532)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer.The investigation of the reported event is currently underway.(expiry date: 09/2021).
 
Event or Problem Description
It was reported that during a stent placement procedure, the stent allegedly partially deployed.It was further reported that stent get caught and fractured while removing.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned for evaluation.Therefore, the investigation is inconclusive for the reported issues.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instruction for use state: 'if excessive force is felt during stent deployment, do not force the stent system.Remove the stent system as possible, and replace with a new unit.' holding and handling of the system throughout deployment was found sufficiently described; in particular the instruction for use state: 'confirm that the introducer sheath is secure and will not move during deployment.To ensure the most accurate placement, firmly hold the black system stability sheath throughout deployment.' under materials required the instruction for use state: 6f (2.0 mm) or larger introducer sheath, 0.035¿ diameter guidewire in addition the packaging pictograms indicate the use of a 6f introducer and a 0.035" guidewire.In regards to pre pta the instructions for use state: 'predilation of the lesion should be performed using standard techniques.' h10: d4 (expiry date: 09/2021).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during a stent placement procedure, the stent allegedly partially deployed.It was further reported that stent get caught and fractured while removing.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT VASCULAR STENT
Common Device Name
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12426169
Report Number9681442-2021-00485
Device Sequence Number3602067
Product Code NIP
UDI-Device Identifier04049519001326
UDI-Public(01)04049519001326
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
P070014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEX071503CS
Device Catalogue NumberEX071503CS
Device Lot NumberANDW4318
Was Device Available for Evaluation? No
Date Returned to Manufacturer04/15/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/10/2021
Supplement Date Received by Manufacturer12/02/2021
Initial Report FDA Received Date09/04/2021
Supplement Report FDA Received Date12/04/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-