BARD PERIPHERAL VASCULAR, INC. ENCOR PROBE; BIOPSY INSTRUMENT Back to Search Results
Model Number ECP017GV
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As this malfunction is considered one event identified by the dealer; only one mdr report will be submitted for the reported quantity affected of nine for this event.Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.One electronic photo was provided and reviewed.The photo shows nine encor probe packaged boxes, crack was noted in all six packages.Based on the photo review, the investigation is confirmed for the reported tear, rip or hole in device packaging (crack and sterile breach on package), as the crack was noted in all nine packages.A definitive root cause for the alleged tear, rip or hole in device packaging could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 04/2022).
 
Event or Problem Description
It was reported that prior to a biopsy procedure, the device package was allegedly damaged.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR PROBE
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key12426231
Report Number2020394-2021-01578
Device Sequence Number19666323
Product Code KNW
UDI-Device Identifier00801741086298
UDI-Public(01)00801741086298
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051158
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberECP017GV
Device Catalogue NumberECP017GV
Device Lot NumberVTER0231
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/09/2021
Initial Report FDA Received Date09/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-