• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 32MM STD LFIT V40 HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 32MM STD LFIT V40 HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 6260-9-132
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Pain (1994); Metal Related Pathology (4530)
Date of Event 05/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding altr , fretting and abnormal ion levels involving a metal head was reported.The event was confirmed through clinician review of the provided medial records.Method & results: product evaluation and results: visual inspection, dimensional, functional inspection, and material analysis were not performed as the item was not returned.Clinician review: a review of the provided medical records by a clinical consultant indicated: revision surgery for fluid collection around left tha and pseudo tumor with mildly elevated co level of 1.9 (0-0.9), and a finding fretting.A revision surgery did occur for a "pseudo tumor".There was a report of a metal reaction and trunnionosis as well as a claim that the lfit head was recalled.That said, there was prior replacement of the head which could have caused changes on the trunnion, there was no metal staining of the tissues; and the co levels were relatively low.The patient had ongoing pain in the hip that was not clearly related to the "pseudo-tumor" and there were no abnormal soft tissue changes noted.The fluid collection was noted on a ct scan and no mars mri data was obtained preoperatively.The pathology results of the tissue nor the explants were not provided for review.The claim of a recalled device would need to be confirmed via the lot numbers.Device history review: indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there has been no other event for the lot referenced.Conclusion: revision surgery for fluid collection around left tha and pseudo tumor with mildly elevated co level of 1.9 (0-0.9), and a finding fretting.A revision surgery did occur for a "pseudo tumor".There was a report of a metal reaction and trunnionosis as well as a claim that the lfit head was recalled.That said, there was prior replacement of the head which could have caused changes on the trunnion, there was no metal staining of the tissues; and the co levels were relatively low.The patient had ongoing pain in the hip that was not clearly related to the "pseudo-tumor" and there were no abnormal soft tissue changes noted.The fluid collection was noted on a ct scan and no mars mri data was obtained preoperatively.The pathology results of the tissue nor the explants were not provided for review.The claim of a recalled device would need to be confirmed via the lot numbers.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
The patient reported she had a left tha on (b)(6) 2014.The patient was revised (b)(6) 2016 due to the screws going into the tendons.The patient has been experiencing pain since surgery.The patient went to the er on (b)(6) 2021 had a ct scan and was told she had lobular fluid surrounding her left hip and was told to follow up with her physician.The patient's physician sent her for blood work which was done on (b)(6) 2021 which showed high levels of chromium.Patient was revised on (b)(6) 2021.Patient is seeking compensation.This pi is for revision 1.The patient's physician sent her for blood work which was done on (b)(6) 2021 which showed high levels of chromium.Patient was revised on (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32MM STD LFIT V40 HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key12432508
Report Number0002249697-2021-01502
Device Sequence Number3503919
Product Code JDI
UDI-Device Identifier07613327013047
UDI-Public07613327013047
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2021
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model Number6260-9-132
Device Catalogue Number6260-9-132
Device Lot Number53986101
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/12/2021
Initial Report FDA Received Date09/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
-
-