Model Number ENSITE-SYSTEM-01 |
Device Problems
Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/16/2021 |
Event Type
malfunction
|
Event Description
|
During a pulmonary vein isolation (pvi) procedure, the hdg catheter interrupted the voxel recording for no apparent reason.After switching to navx mode, all catheters jumped up and down about a centimeter in an uncontrolled manner.Since the problem could not be resolved, the procedure was aborted.The patient was stable before and after the examination.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.In addition, no log files were received.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined. .
|
|