• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ENSITE X EP SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ENSITE X EP SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number ENSITE-SYSTEM-01
Device Problems Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2021
Event Type  malfunction  
Event Description
During a pulmonary vein isolation (pvi) procedure, the hdg catheter interrupted the voxel recording for no apparent reason.After switching to navx mode, all catheters jumped up and down about a centimeter in an uncontrolled manner.Since the problem could not be resolved, the procedure was aborted.The patient was stable before and after the examination.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.In addition, no log files were received.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE X EP SYSTEM
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
MDR Report Key12432739
MDR Text Key270085467
Report Number2184149-2021-00259
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberENSITE-SYSTEM-01
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/16/2021
Initial Date FDA Received09/07/2021
Supplement Dates Manufacturer Received09/21/2021
Supplement Dates FDA Received09/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-