LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 10-88-00 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).The affected device was returned to the manufacturer site.The reported issue was confirmed and blood residues were detected in the pump and on the two circuit boards.In addition, the motor connector was found to be not properly engaged on the circuit board and the pump head was anodized and scratched.The sealing, the pump head and two other components were replaced to solve the problem.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that s5 mast roller pump gave a motor control failure error message during priming.There was no patient involvement.
|
|
Manufacturer Narrative
|
H.10: the root cause of the reported event could be traced back to a malfunction of the motor control board and motor amplifier board, due to a significant blood spill into the pump head, affecting the boards functionality.
|
|
Event Description
|
See initial report.
|
|