| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the impactor fractured during a knee procedure.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Updated: b4, g3, g6, h2, h3, h6, and h10.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned instrument identified signs of repeated use (nicked/gouged) and that it is fractured.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment of the reported event, it was determined an mdr should not have been filed under the current mfr number.This event will be reported on 0001825034-2022-00495.
|
| |
|
Event or Problem Description
|
|
Upon reassessment of the reported event, it was determined an mdr should not have been filed under the current mfr number.This event will be reported on 0001825034-2022-00495.
|
| |
|