Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the impactor fractured during a knee procedure.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Updated: b4, g3, g6, h2, h3, h6, and h10.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned instrument identified signs of repeated use (nicked/gouged) and that it is fractured.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
Upon reassessment of the reported event, it was determined an mdr should not have been filed under the current mfr number.This event will be reported on 0001825034-2022-00495.
 
Event or Problem Description
Upon reassessment of the reported event, it was determined an mdr should not have been filed under the current mfr number.This event will be reported on 0001825034-2022-00495.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING FEMORAL IMPACTOR PAD
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12433303
Report Number0001822565-2021-02477
Device Sequence Number17415190
Product Code HWA
UDI-Device Identifier00880304809352
UDI-Public(01)00880304809352(10)63713419
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42509909000
Device Lot Number63713419
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2021
Supplement Date Received by Manufacturer09/20/2021
03/03/2022
Initial Report FDA Received Date09/07/2021
Supplement Report FDA Received Date09/22/2021
03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexMale
-
-