• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER Back to Search Results
Model Number 0166SI16
Medical Device Problem Code Partial Blockage (1065)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that for the last nine months, the catheters had been getting stopped up and the tip eyelets were crusted with a gritty sandy substance almost like it was disintegrating while in use.It was stated that they were alternating with gentamycin and acidic acid for flushing and they could not flush the catheter.It was also stated that three catheters would not deflate after use.It was noted that the patient had no medical condition identified that would cause this type of issue.
 
Event or Problem Description
It was reported that for the last nine months, the catheters had been getting stopped up and the tip eyelets were crusted with a gritty sandy substance almost like it was disintegrating while in use.It was stated that they were alternating with gentamycin and acidic acid for flushing and they could not flush the catheter.It was also stated that three catheters would not deflate after use.It was noted that the patient had no medical condition identified that would cause this type of issue.
 
Additional Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "instructions for use: 1.Occlude drainage tubing a minimum of 3 inches below the sampling port by kinking the tubing until urine is visible under the access site.2.Swab surface of bard® ez-lok® sampling port with antiseptic wipe.3.Using aseptic technique, position the syringe in the center of the sampling port.The syringe should be held perpendicular to the surface of the sampling port (at approximately 80 - 100 degree angle).Press the syringe firmly and twist gently to access the sampling port.4.Slowly aspirate urine sample into syringe and remove syringe from sample port.5.Unkink tubing and transfer urine specimen into specimen cup or follow hospital protocol.Discard syringe according to hospital protocol.6.Follow established hospital protocol for specimen labeling and transport to lab.Sterile: contents of inner wrap are sterile unless package has been opened or damaged.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not re-sterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Urology urology cdc guidelines for appropriate indications for indwelling urethral catheter use ¿ patient has acute urinary retention or bladder outlet obstruction ¿ need for accurate urine output measurements ¿ use for selected surgical procedures ¿ to assist in healing of open sacral or perineal wounds ¿ patient requires prolonged immobilization ¿ to improve comfort for end of life care proper techniques for urinary catheter insertion perform hand hygiene immediately before and after insertion insert urinary catheters using aseptic technique and sterile equipment.Use the smallest foley catheter possible, consistent with good drainage document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in the patient record.Proper techniques for urinary catheter maintenance secure the foley catheter.Use the statlock® foley stabilization device if provided maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions.Maintain unobstructed urine flow and keep the catheter and collection tube free from kinking.Keep the collection bag below the level of the bladder or hips at all times.Empty the collection bag regularly using a separate, clean collection container for each patient directions for use/patient education information insert foley catheters only for appropriate indications and leave in place only as long as needed this catheter is intended for use in the drainage and/or collection and/or measurement of urine.* * generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder.However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as nephrostomy tract." the actual/suspected device was evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12433847
Report Number1018233-2021-05460
Device Sequence Number3391570
Product Code MJC
UDI-Device Identifier00801741016844
UDI-Public(01)00801741016844
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K040658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2025
Device Model Number0166SI16
Device Catalogue Number0166SI16
Device Lot NumberMYEU4766
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/14/2021
Supplement Date Received by Manufacturer10/22/2021
Initial Report FDA Received Date09/07/2021
Supplement Report FDA Received Date11/11/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-