According to the reporter, during use, it was noted that blood leak (hole/damage) was found on the extension tube near the luer adap ter(blue).The catheter was not repaired and the tube was replaced to resolve the issue.There was nothing unusual observed with the product prior to use and there were no other visible defects/damages found on the product.It was also mentioned that flushing was done prior to use and had a normal result.No other product was used with the device.It was also stated that the clamp was moved periodically and a metal clamp was not used during insertion.The cleaning agent(s) was not mixed and sepsiderm was not used to clean the device.The patient did not used any type of cleaning agent/antibiotic on the catheter.There were no protocol changed to the recently used cleaners.There was no blood loss and blood transfusion was not needed.It was also stated that the medical intervention done was to remove the catheter and replaced with a new device.The procedure was completed.Tego was not utilized and there was no luer adapter issue.There was no reported patient injury.
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted there was a cut in the extension tube which led to a leak.It was reported that there was a leak on the extension tube.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur when contact is made with a sharp surgical instrument during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: inspect the catheter frequently for nicks, scrapes, cuts, etc.Which could impair performance.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|