| Catalog Number |
0026880 |
| Medical Device Problem Codes |
Corroded (1131); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As reported, fluid leakage was noted inside of the hydrosurg plus battery compartment at the end of the case.The subject product was returned for evaluation.Visual examination identified corrosion of the internal components, which was due to fluid ingress into the motor housing.Cracks and excessive rtv were identified on the motor mount allowing fluid to pass through the motor mount and down the sides of the motor to the pc board and battery compartment.Based on the sample evaluation and investigation performed, the root cause for the issues experience by the user are determined to be manufacturing related.
|
| |
|
Event or Problem Description
|
|
As reported, during an unknown procedure on (b)(6) 2021, a bard/davol hydrosurg irrigation device was used.It was reported that at the end of the case while removing the irrigator from the iv pole/neptune, the irrigator was leaking fluid from the battery pack.There was no performance issue and the device functioned as intended.There was no reported patient or user harm.
|
| |
|