Catalog Number 7N8371
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an one-link non-dehp y-type microbore catheter extension set had a broken y-connector (the tubing separated from the male luer).The issue was identified during patient use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed and observed a male luer lock is apart from the tubing.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SETS WITH ONE-LINK NEEDLE-FREE IV CONNECTOR
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key12434949
Report Number1416980-2021-05479
Device Sequence Number11618211
Product Code FPA
UDI-Device Identifier00085412091822
UDI-Public(01)00085412091822
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
K132734
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number7N8371
Device Lot NumberDR20K12017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2021
Supplement Date Received by Manufacturer11/03/2021
Initial Report FDA Received Date09/07/2021
Supplement Report FDA Received Date11/08/2021
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PUMP; PUMP
Patient SexUnknown
-
-