| Model Number |
TA6035L |
| Medical Device Problem Codes |
Component Missing (2306); Mechanics Altered (2984)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Decreased Peak Expiratory Flow rate (2435); Unintended Radiation Exposure (4565)
|
| Date of Event |
08/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during right medial liver lobectomy, as liver lobe was being transacted away from the stapler, after cutting the tissue, it was discovered that the device was unable to fire.The tissue started to pull away and signification bleeding of 800cc occurred.The dog had arrest within 5-10 seconds of bleeding.Surgical time was extended by at least 30 minutes.Resuscitation efforts were unsuccessful and the patient died.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the instrument was received with a fully applied 3.5mm single use loading unit (sulu).Microscopic evaluation of the sulu noted some pusher fingers were damaged indicating the device was applied over an obstruction.Visual inspection of the sealed product noted no abnormalities.It was reported that the staples did not deploy from the stapler after the handle was squeezed.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: there was damage on the pushers of the loading unit, consistent to application over an obstruction.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during veterinary right medial liver lobectomy, as liver lobe was being transacted away from the stapler, after cutting the tissue, it was discovered that the device was unable to fire.The tissue started to pull away and signification bleeding of 800cc occurred.The dog had arrest within 5-10 seconds of bleeding.Surgical time was extended by at least 30 minutes.Resuscitation efforts were unsuccessful and the patient died.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during veterinary right medial liver lobectomy, as liver lobe was being transacted away from the stapler, after cutting the tissue, it was discovered that the device was unable to fire.There were no staples on the tissue nor the cartridge when inspected visually and radiographicallly.The tissue started to pull away and signification bleeding of 800cc occurred.The dog had arrest within 5-10 seconds of bleeding.Surgical time was extended by at least 30 minutes.Resuscitation efforts were unsuccessful and the patient died.There were no staples on the tissue nor the cartridge when inspected visually and radiographicallly.
|
| |
|