| Model Number |
150819 |
| Medical Device Problem Code |
Restricted Flow rate (1248)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the urine was backing up.The representative did troubleshooting, asked about the position of the hose, and noted it to be straight.Also, asked if the patient had blown on the bag or added water to separate the inside and provided a few tips.The patient had been using the product for less than 90 days.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive as no sample returned for evaluation.A potential root cause for this failure mode could be due to obstruction in tube.It was unknown whether the device had met relevant specifications.The product was used for the treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number was unknown therefore the device history record could not be reviewed.The instructions for use were found adequate and state the following: "directions for use: 1.Separate notches within circles.Pull straps through holes and around leg.2.Position bag on leg with flutter valve at top.3.Attach catheter or extension tubing to top inlet.When wearing bag below knee, attach bard® extension tubing.When using extension tubing, make sure to connect tube at least to the 3rd taper of the connector.4.To empty dispoz-a-bag®, push green lever on flip-flo¿ valve out and down.Important: be sure to reclose flip-flo¿ valve after emptying bag.5.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable, local, state and federal laws and regulations." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the urine was backing up.The representative did troubleshooting, asked about the position of the hose and noted it to be straight.Also asked if the patient had blown on the bag or added water to separate the inside and provided a few tips.The patient had been using the product for less than 90 days.
|
| |
|