| Catalog Number |
195-160 |
| Medical Device Problem Codes |
False Negative Result (1225); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.A supplemental report will be provided after completion.
|
| |
|
Event or Problem Description
|
|
The consumer reported a unconfirmed false negative result with the binaxnow¿ covid-19 antigen self test performed on (b)(6) 2021.Repeat testing was performed and generated a positive result.Consumer stated he was not feeling okay.No additional patient information, including treatment and outcome, was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot lot 144958 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 195-160 / lot 144958 , test base part number 195-430h / lot 140390.The lot met the required release specifications.A review of the complaints reported as false positive and false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 144958 showed that the complaint rate is (b)(4).Abbott diagnostics scarborough was unable to determine the exact root cause of the reported issue.However a possible assignable root cause is sample interference or cross contamination.
|
| |
|
Event or Problem Description
|
|
The consumer reported a conflicting result with the binaxnow¿ covid-19 antigen self test performed on 16aug2021.Consumer stated he was not feeling okay.No additional patient information, including treatment and outcome, was provided.
|
| |
|