| Model Number |
1012-04-010 |
| Medical Device Problem Codes |
Osseointegration Problem (3003); Migration (4003)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646)
|
| Date of Event |
01/27/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Compass case (b)(4) report: it was reported that the stem subsided.Implant loosening was also noted of the trilock component-noncemented.The sale rep was contacted by the surgeon that she had a trilock implanted some time ago that had subsided in one of her patients.She did say that the patient missed their post-op office visit.She showed back up with a painful hip.She explained the stem had subsided and she wanted to revise it to a stryker and have liners available for exchange for the pinnacle cup.The pinnacle liner was exchanged and the stem and head was revised to stryker.The ref: (b)(4) lot: 39030g pinnacle altrx liner 36mm/52mm was used to replace the old liner.The 52mm pinnacle porocoat shell ref:(b)(4) lot:c43763 was left in patient.Doi: (b)(6) 2016; dor: (b)(6) 2021; affected side: right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|